A 32-Channel 85.4 dB SNDR Time-Multiplexed Neural Recording Front-End Achieving Within-Conversion Artifact Recovery

A 32-Channel 85.4 dB SNDR Time-Multiplexed Neural Recording Front-End Achieving Within-Conversion Artifact Recovery 150 150

Abstract:

This article presents a 32-channel time-multiplexed highly digital neural recording front-end (RFE) that exhibits sub- $8.8~\mu $ s recovery latency from large stimulation artifacts while delivering high-resolution data at the Nyquist rate. The RFE is time-shared across 32 channels for low-frequency electrocorticography (ECoG) recording and across four channels for high-frequency action potentials (…

View on IEEE Xplore

A 10.2 V 8-Channel Neural Stimulator With Nearly Constant Efficiency Across 90% Current Range and a TDM ADC-Based One-Shot Charge Balancing With < 6 mV Residue

A 10.2 V 8-Channel Neural Stimulator With Nearly Constant Efficiency Across 90% Current Range and a TDM ADC-Based One-Shot Charge Balancing With < 6 mV Residue 150 150

Abstract:

This letter presents an 8-channel current-mode stimulator achieving a ±10.2 V voltage compliance in a standard CMOS process, supporting up to 4.5 mA stimulation current per channel. The proposed dynamic charge pump (DCP), which adaptively sizes its switches based on the stimulation current, helps achieve ~15% higher power efficiency at low currents compared …